Search for a product

Read catalog
Customer support > Maintenance

Maintenance

.
CFR & ULTRA
Cooling system: 

Water : Distilled water with 1MΩ.cm to 5MΩ.cm resistivity or 50/50 Reagent Grade Ethylene Glycol/distilled water mixture if ordered with the Ethylene Glycol option

DI cartridge exchange : Every 6 months



Environmental conditions: 

Ambiant temperature range : 10°C to 40°C

Storage temperature range : 5°C to 50°C

 

For additional information, please consult the user's manual or contact customerservice@lumibird.com

EVERGREEN
Cooling system: 

Water : Distilled water with 1MΩ.cm to 5MΩ.cm resistivity

DI cartridge exchange : Every 6 months



Environmental conditions: 

Ambiant temperature range : 18°C to 28°C

Storage temperature range : 5°C to 50°C

For additional information, please consult the user's manual or contact customerservice@lumibird.com

Q-SMART 450/850
Cooling system:

Water : Distilled/Demineralized water with 1MΩ.cm to 5MΩ.cm resistivity

DI cartridge exchange : Every 6 months

Environmental conditions :

Ambiant temperature range : 18°C to 28°C

Storage temperature range : 5°C to 50°C

 

For additional information, please consult the user's manual or contact customerservice@lumibird.com